8000R

Availability: In Stock
Reusable Reflectance Pulse Oximetry Sensor

Nonin 8000R reflectance pulse oximetry sensors provide a convenient application site on the forehead for stress testing and continuous monitoring where digits are not the preferred location. With Nonin PureLight® sensor technology and PureSAT® signal processing technology, you can be sure you’re getting consistent results from patient to patient and sensor to sensor. The product consists of the 8000R sensor, 8000H plastic applicator caps, and adhesive rings.

Category:

SpO2 Accuracy 1, 2:

Range Oxygen Saturation (Arms*)
70 – 100% ±2
70 – 80% ±3
80 – 90% ±2
90 – 100% ±2

SpO2 Low Perfusion Accuracy: 70 to 100% ±2 digits (Arms*)1
Pulse Rate Accuracy: 40 to 240 BPM ±3 digits (Arms*)1
Pulse Rate Low Perfusion Accuracy: 40 to 240 BPM ±3 digits (Arms*)1
Temperature: 3, 4
Operating: -20 °C to 50 °C (-4 °F to 122 °F)
Storage/Transportation: -40 °C to 70 °C (-40 °F to 158 °F)

Humidity: 3, 4
Operating: 10% to 95% non-condensing
Storage/Transportation: 10% to 95% non-condensing

* ±1 Arms encompasses 68% of the population at zero bias.
1 SpO2 accuracy was conducted during induced hypoxia study on healthy subjects over the range of 70 % to 100 %.
2 Accuracy specifications based on Nonin’s PureSAT® SpO2 technology and PureLight® sensor technology.
3 Additional accuracy and performance information can be found in the sensor accuracy document on the operator’s manual CD.
4 For combined oximeter/sensor specifications, refer to the applicable oximetry system’s operator’s manual.
5 Range as tested with Nonin’s PureSAT SpO2 technology.

Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 3 mW nominal
Infrared: 910 nanometers @ 3 mW nominal
** This information is especially useful for clinicians performing photodynamic therapy.

Compliance
This product complies with ISO 10993.
Not made from natural rubber latex.

Documentation

Videos

8000R Reflectance Forehead Pulse Oximeter Sensor

Resources