SenSmart 8006 Hybrid Sensor

Availability: In Stock
For use with the SenSmart X-100 Universal Oximetry System

Regional oximetry (rSO2) monitoring helps clinicians identify desaturations during critical procedures. The SenSmart® Model X-100 Universal Oximetry System provides reliable rSO2 and SpO2 measurements they can use right away.

The SenSmart® 8006 Hybrid Sensor provides reliable rSO2 readings with a reusable sensor and a disposable patient interface. It reduces cost per procedure, helping make cerebral oximetry monitoring and its benefits available to more patients.

Category:

rSO2 Accuracy (Arms*): 50% to 100% rSO2

Accuracy Right Left Both Hypercapnia Hypocapnia
Absolute 4.1 3.8 3.9 5.1 3.3
Trend 1.9 3.0 2.5 3.4 3.8

rSO2 accuracy testing was conducted during induced hypoxia studies on male and female, over 18 years of age, healthy, light- to dark-skinned subjects in an independent research laboratory. The measured regional hemoglobin saturation value (rSO2) of the sensors was compared to arterial/venous hemoglobin oxygen (SavO2) value, determined from venous and arterial blood samples. The model used for blood in the brain was 70% venous and 30% arterial, which is applicable under normocapnic conditions. The venous blood was drawn from the right jugular bulb. The accuracy of the sensors in comparison to the blood gas analyzer samples measured over the rSO2 range of 45–100%. Accuracy data was calculated using the root-meansquared (Arms value) for all subjects, per ISO 80601-2-61:2011, Medical Electrical Equipment—Particular requirements for basic safety and essential performance of pulse oximeter equipment.

Inter/Intra Sensor Repeatability Accuracy: ±2 digits (Arms*)
Temperature:
Operating: -5 °C to 40 °C (23 °F to 104 °F)
Storage/Transportation: -30 °C to 70 °C (-22 °F to 158 °F)
Device temperature will not exceed 41 °C as measured during a controlled environment test.

Humidity:
Operating: 10 % to 90 % non-condensing
Storage/Transportation: 10 % to 95 % non-condensing
* ±1 Arms encompasses 68% of the population at zero bias.

Magnetic Field Strength (8006PI):
Less than 1 gauss (0.1T) at a distance of 12cm.

Measurement Wavelengths and Output Power**
730 nanometers @ 3.0 mW maximum average power
760 nanometers @ 4.5 mW maximum average power
810 nanometers @ 3.2 mW maximum average power
880 nanometers @ 4.5 mW maximum average power
** This information is especially useful for clinicians performing photodynamic therapy.

Compliance
This product complies with ISO 10993-1.